Rushing the shut down procedure can cause errors on Spin SR

We regularly get a calibration error form the Marzhauser stage when running the Spin SR10 system. We tried to tackle this problem in many ways and did solve it yet. Please be sure to notify us when you encounter the issue. One cause could be that users shut down the system too fast (when settings…

Read more

Updated Gas Supply CellVivo

We have replaced the CO2 sensor on the gas mixer of the CellVivo System. It is working properly again. ALWAYS make sure the valve for gas supply is opened when you start. And make sure the valve is closed again when you end your live cell imaging session.

Read more

DEMO Microscopes from 17th-22nd of May

We are testing different microscopes to replace the Leica SPE2 “workhorse” that is having frequent technical failures and is heavily overbooked nowadays. Our aim is to increase the imaging capacity at the CCI by improving the workflow to do your imaging. We sent a mail with a request to give us samples you would like…

Read more

Leica SPE2 40x HCX oil objective defect

The 40x oil on the Leica SPE-II CLSM showed severe chromatic aberration in Z for UV fluorescence. The objective is removed from the system and its aberration is being validated on another Leica machine in the UU campus. We probably need to repair or replace it. Our objectives are very costly and vulnerable, please handle…

Read more

Update NIKON A1R upcoming

We are grateful to the Veterinary Faculty Utrecht to support an upgrade for the NIKON  A1R system. Many users will benefit from this upgrade since it will vastly improve confocal imaging speeds. The upgrade will be performed  around the last week of May 2019.

Read more

Soledad appointed, Cryostate moved and supported by Judith in NG O.53

Soledad Ordoñez-Alvarez has been appointed for 2 days a week as a new staff member of the CCI. She specializes in instructing new users and adds expertise in image analyses to the CCI. Unfortunately, Rob Bleumink had to retire. Since we miss his expertise in cryosectioning (as for countless other issues), we moved the cryostate…

Read more

OMX service performed CAM4 currently not functional

The OMX has been serviced , the OTFs have been updated and saved as default for SIM reconstructions. A 10x objective is available now to perform tile scans for sample overviews. The 4th camera is non functional due to a power failure, repair parts are on order. So for now, the far red detection is…

Read more