We have issues with the CO2 dispense system from CellVivo again. Currently an old TOKAI hit CO2 supply runs on the Evident Spin SR10 system. Turn on by the “on” button and make sure the gas supply valve is open as normally. The system is hard set to mix 5% CO2 with ambient air, check…
We are pleased to announce the purchase of a new confocal scanning microscope. The plan is that the machine will be able to take over part of the workload on the busy Evident Spinning disk. We chose to get functionality on the system to flexibly automate imaging, with a focus on being able to do…
We regularly get a calibration error form the Marzhauser stage when running the Spin SR10 system. We tried to tackle this problem in many ways and did solve it yet. Please be sure to notify us when you encounter the issue. One cause could be that users shut down the system too fast (when settings…
We have replaced the CO2 sensor on the gas mixer of the CellVivo System. It is working properly again. ALWAYS make sure the valve for gas supply is opened when you start. And make sure the valve is closed again when you end your live cell imaging session.